Safety test METRISO PRIME+
METRISO PRIME+ digital high-voltage insulation tester, test voltage up to 5000V DC, voltage measurement to 1000V to detect the residual voltage of the device under test, up to 1000V for capacitive devices to be tested, such as cables and coils, may be due to The test voltage is used to charge it, and the meter can be used to discharge it. The item can show the voltage value drop. The measurement process is in accordance with EN61557. The test voltage of DIN VDE 0413 is 100, 250, 500 or 1000V. The rated current is 1Ma.
METRISO PRIME+
Digital high-voltage insulation tester, test voltage up to 5000V DC, voltage measurement to 1000V
Detect the residual voltage of the device under test, up to 1000V
Discharge the capacitive device under test
Capacitive devices under test, such as cables and coils, may be charged by the test voltage, which can be discharged by the meter. The item can show the voltage value drop process.
Measurement according to EN61557, DIN VDE 0413
Test voltage is 100, 250, 500 or 1000V, rated current is 1Ma
High insulation performance measuring cable
For safety reasons, the high insulation performance measuring cable is fixedly connected to the tester sampling to avoid the risk of inadvertently interrupting the cable connection.
Measuring Polarization Index (PI)
Electrical machines are usually subjected to polarization index testing. The temperature testing angle is an extension of the insulation resistance test. The DC measurement provided by METRISOPRIME+, the voltage can last for ten minutes in the insulation test. The measurement results were recorded after one minute of testing and one measurement was recorded ten minutes later. If the measured value at ten minutes is greater than the measured value at one minute, say
The insulation is good. The ratio of the two measurements is the polarization index. During the insulation test, the carrier inside the chargeable material is aligned due to the long measurement voltage, resulting in polarization. The polarization index indicates whether the internal carriers of the charged insulating material can also move, that is, whether they can also be polarized, that is, the polarization index indicates the state of the insulating property.
Previous: Insulation layer test device for bendable current sensor
下一条: Insulation layer test device for bendable current sensor