HFCN Test Fixture ECG test
Verify that the electrosurgical knife interferes with the test fixture of the ECG device or monitor. The dual modular design meets the test requirements of China YY, IEC and US EC, and complies with the latest international and domestic standards: IEC60601-2-25, IEC60601- 2-27, YY1079, AAMI EC13 Overview: ECG tester professional test fixture to improve test convenience and safety dual module design, in line with the latest international and domestic standards: IEC60601-2-25, IEC60601-2-27, YY1079 AAMI EC13
Verify that the electrosurgical knife interferes with the test fixture of the ECG device or monitor. The dual modular design meets the test requirements of China YY, IEC and US EC, and complies with the latest international and domestic standards: IEC60601-2-25, IEC60601- 2-27, YY1079, AAMI EC13
Overview:
Professional test fixtures improve test convenience and safety
Dual module design, in line with multinational standards
Built-in 51k/47nF simulated skin impedance
1mV ECG signal required to simulate electrosurgical interference suppression (ESIS) testing
Can withstand electric knife power >300W
Built-in non-inductive value (<5μh), high="" frequency="" power="">5μh),>
Sharing junction box with SECG 4.0 and MECG 2.0 testers
Previous: SECG standard auxiliary software
下一条: ECG test CMRR 3.0